Conference Proceeding

Artefacts Removal from Optical Coherence Tomography Angiography

Authors: , , , ,

Abstract

This paper presents a new approach for artefacts removal from optical coherence tomography angiography (OCTA). The artefacts mainly arise as a result of distortion due to eye movements during OCT scanning process. These distortions manifest themselves as visible motion artefacts when doctors review the enface image of OCTA data. To remove these artefacts, firstly we perform motion registration for the captured OCT volume data and subsequently perform motion correction to obtain the registered OCT data. Next, we compute the OCTA from the registered OCT data using an enhanced correlation mapping technique. Thereafter, we compute the enface image from the OCTA data. In the next step, we attempt to locate regions where there is misalignment in the OCT frames of the various B-scans. Finally, we attempt to restore the regions where correct data is postulated to be absent. Our experimental results demonstrate the effectiveness of our proposed approach.

Keywords:

How to Cite: Ong, E. P. , Cheng, J. , Quan, Y. , Xu, G. & Wong, D. W. (2016) “Artefacts Removal from Optical Coherence Tomography Angiography”, Proceedings of the Ophthalmic Medical Image Analysis International Workshop. 3(2016). doi: https://doi.org/10.17077/omia.1049